| Model Number |
0015790 |
| Medical Device Problem Code |
Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/21/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged while attempting to load the patient, the crew was unable to retract the undercarriage.The patient was transferred to another cot and transported.No injuries were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The cot was evaluated by an authorized field technician at the complainant's site.A visual and functional evaluation was conducted and there was no observation of any malfunction.The cot was functioning according to specification and the reported incident could not be duplicated.It was observed the crew was operating the release system incorrectly.This type of cot is new to the service and the release system is operated differently than their prior product.While onsite, the technician showed the crew the proper loading technique for this cot.The complainant did advise the cot was able to be operated correctly after the reported incident and it was never removed from service.The ifu for the cot provides sufficient instructions on how to operate the release mechanism to raise the undercarriage during the loading process.
|
| |
|
Event or Problem Description
|
|
Complainant alleged while attempting to load the patient, the crew was unable to retract the undercarriage.The patient was transferred to another cot and transported.No injuries were reported.
|
| |
|