| Model Number |
24970A |
| Medical Device Problem Code |
Computer Operating System Problem (2898)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/27/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the programmer was freezing up during use.The programmer was returned for service.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: manufacturer's analysis confirmed the customer comment that the programmer was freezing up during use.It appeared that the base station analyzer die-stack experienced a power on reset that caused an analyzer electrical reset error.Use conditions were identified that are known to cause this issue.It was likely the unit experienced a power on reset, due to electrocautery while the base was sitting on a metal surface.The prorgrammer was returned to service.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|