| Model Number |
537540 |
| Medical Device Problem Code |
Incorrect Or Inadequate Test Results (2456)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer in the united states notified biomérieux of an "odd result" associated with etest® ceftaroline cpt 32 us b30 (reference 537540).The issue came up during testing of a quality control sample.Another test from a different lot was opened and the test run appropriately.There is no patient involved this event occurred during a quality control test.An internal biomérieux investigation will be initiated.
|
| |
|
Additional Manufacturer Narrative
|
|
This investigation was initiated due to an odd result with etest ceftaroline cpt 32 us b30 ref 537540 and staphylococcus aureus atcc 29213.Given that the customer could not provide biomerieux with the lot numbers he used, the batches sent to this customer were searched internally.In conclusion, no batches of this reference 537540 have been sent to this customer since january 2014.The shelf life of this reference being two years, it's impossible for the customer to have used a non-expired lot.Because of this, no further investigation will be performed.
|
| |
|
Search Alerts/Recalls
|