MAUDE Adverse Event Report: ORTHO-CLINICAL DIAGNOSTICS VITROS CHEMISTRY PRODUCTS CALIBRATOR KIT 2; IN VITRO DIAGNOSTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ORTHO-CLINICAL DIAGNOSTICS VITROS CHEMISTRY PRODUCTS CALIBRATOR KIT 2; IN VITRO DIAGNOSTICS Back to Search Results
Catalog Number 1662659
Device Problems High Test Results (2457); Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/08/2017
Event Type  malfunction  
Manufacturer Narrative
The investigation determined that higher and lower than expected vitros na+ quality control results were obtained when processing non-vitros biorad lot 31830 controls and vitros performance verifier control lot l5005 on two vitros 5600 integrated systems.The investigation determined the assignable cause to be user error due to a combination of not calibrating na+ slides for an electrolyte reference fluid (erf) lot change and improper calibrator kit fluid preparation.The lower than expected biorad level 2 na+ result was obtained using a new vitros erf lot which was not the erf lot used during the initial na+ calibration event.In the ¿when to calibrate¿ section of the vitros na+ instructions for use (ifu) the customer is instructed to calibrate na+ whenever the vitros electrolyte reference fluid lot number changes.In response to the low biorad quality control results, the customer recalibrated vitros na+ multiple times using 2 different calibrator kit lots (lots 257 and 226) and obtained higher than expected quality control results on biorad and vitros performance verifier controls.It was concluded that the calibrator fluids were not prepared correctly which subsequently caused a suboptimal na+ calibration and the higher than expected biorad qc results.Acceptable vitros na+ quality control results were obtained after a recalibration event was performed using properly handled calibrator fluids and the new erf lot.The investigation found no indication the vitros 5600 chemistry system or vitros na+ slide lot contributed to the event.
 
Event Description
The investigation determined that higher and lower than expected vitros na+ quality control results were obtained when processing non-vitros biorad controls and vitros performance verifier controls on two vitros 5600 integrated systems.Vitros 5600 integrated system (s/n (b)(4)): (b)(6).Vitros 5600 integrated system (s/n (b)(4)): biorad lot 31831 level 1 = 139.34 and 141.66 mmol/l versus expected 124.3 mmol/l.Biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected on patient samples.It is unknown if patient results were reported during the time frame of the event.Ortho was not made aware of any misreported results and there was no allegation of patient harm as a result of this event.This report is number 1 of 3 mdr¿s for this event.Three (3) 3500a forms are being submitted for this event as 3 devices were involved.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITROS CHEMISTRY PRODUCTS CALIBRATOR KIT 2
Type of Device
IN VITRO DIAGNOSTICS
Manufacturer (Section D)
ORTHO-CLINICAL DIAGNOSTICS
100 indigo creek drive
rochester NY 14626
Manufacturer (Section G)
ORTHO-CLINICAL DIAGNOSTICS
1000 lee road
rochester NY 14606
Manufacturer Contact
james a stevens
100 indigo creek drive
rochester, NY 14626
5854533000
MDR Report Key6921622
MDR Text Key90209789
Report Number1319808-2017-00018
Device Sequence Number1
Product Code JIX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 10/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/19/2018
Device Catalogue Number1662659
Device Lot Number257
Other Device ID Number10758750009503
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/08/2017
Initial Date FDA Received10/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/19/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-