MPRI CAPSUREFIX NOVUS; ELECTRODE, PACEMAKER, PERMANENT Back to Search Results
Model Number 5076-85
Medical Device Problem Codes Device Dislodged or Dislocated (2923); High Capture Threshold (3266)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/23/2017
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the right atrial (ra) lead and right ventricular (rv) leads dislodged and exhibited high thresholds.Both leads were explanted and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the full lead was returned in segments, analyzed, and no anomalies were found.Visual analysis of the lead indicated apparent explant damage.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS
Common Device Name
ELECTRODE, PACEMAKER, PERMANENT
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key6922679
Report Number2649622-2017-11131
Device Sequence Number16356566
Product Code DTB
UDI-Device Identifier00643169550797
UDI-Public00643169550797
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P930039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/06/2017
Device Model Number5076-85
Device Catalogue Number5076-85
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2017
Initial Date Received by Manufacturer 08/23/2017
Supplement Date Received by Manufacturer11/21/2017
Initial Report FDA Received Date10/06/2017
Supplement Report FDA Received Date02/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
A2DR01 IPG, 5076-65 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age69 YR
-
-