| Catalog Number |
RBY4C0305 |
| Medical Device Problem Codes |
Bent (1059); Device Damaged Prior to Use (2284); Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/18/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the ruby coil pusher assembly was kinked approximately 68.0 cm from the proximal end.Conclusions: evaluation of the returned ruby coil confirmed its pusher assembly was kinked.This damage may have occurred due to forceful handling of the ruby coil during attempts to re-sheath.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure using ruby coils.During the procedure, while the hospital technician was removing a ruby coil from the packaging hoop, the coil accidentally became unsheathed.Upon attempting to re-sheath the ruby coil, the pusher wire became bent.The damage to the ruby coil occurred prior to use and therefore, it was not used in the procedure.The procedure was completed using a new ruby coil.
|
| |
|