• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY4C0305
Medical Device Problem Codes Bent (1059); Device Damaged Prior to Use (2284); Device Handling Problem (3265)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the ruby coil pusher assembly was kinked approximately 68.0 cm from the proximal end.Conclusions: evaluation of the returned ruby coil confirmed its pusher assembly was kinked.This damage may have occurred due to forceful handling of the ruby coil during attempts to re-sheath.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using ruby coils.During the procedure, while the hospital technician was removing a ruby coil from the packaging hoop, the coil accidentally became unsheathed.Upon attempting to re-sheath the ruby coil, the pusher wire became bent.The damage to the ruby coil occurred prior to use and therefore, it was not used in the procedure.The procedure was completed using a new ruby coil.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key6925410
Report Number3005168196-2017-01724
Device Sequence Number852274
Product Code HCG
UDI-Device Identifier00814548013145
UDI-Public00814548013145
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Radiologic Technologist
Device Expiration Date06/14/2025
Device Catalogue NumberRBY4C0305
Device Lot NumberF77707
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2017
Initial Report FDA Received Date10/06/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
-
-