|
It was reported by the international customer that, when handling the catheter contained in the wayne pneumothorax set, the joints created between the different components were loosened too easily.The circumstances surrounding the usage and handling of the device are not known.The product is reportedly not available for return.No further information was provided by the customer.
|
|
Investigation ¿ evaluation: a review of the complaint history, device history record, instructions for use (ifu), specifications, and quality control data was conducted during the investigation.The complaint device was not returned therefore, device failure analysis and physical examination of the device used in this case could not be performed.However, a photo was provided and reviewed.The photo appears to indicate a loose or split fitting with the device.If the device is returned for evaluation, a thorough analysis and determination will be done and the record will be updated at that time.Review of production and quality documentation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.A thorough review of the device history record did not observe any non-conformance's that may have contributed to this incident associated with the complaint.There were two other reported complaints for this lot number from the same facility.The device is shipped with the instructions for use that states the proper warnings and precautions.Specifically; ¿device is not recommended for large fluid accumulation or hemothorax.All connections must be secure and airtight.Perform inspections of the catheter and connections regularly.Secure catheter in position at the entry site by using a bio-occlusive dressing or suturing if desired.The catheter and connected drain lines should be secured to the patient.Excessive tension on catheter connections may cause catheter/hub separation or accidental catheter dislodgement." based on the information provided, no product returned, and the results of our investigation; a definitive root cause could not be determined.Per the quality engineering risk assessment, no further action is required.Appropriate personnel have been notified and monitoring will continue to be performed for similar complaints.
|