| Model Number |
GMR40 |
| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Codes |
Abrasion (1689); Bruise/Contusion (1754)
|
| Date of Event |
09/21/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A design modification will be implemented to the detector bin.This system has been taken out of service until this modification is put in place.Additionally, the foot crush label will be added to all systems in the field and/or in inventory.The noted design modification, which will focus on raising the gap distance to a more acceptable level where this injury will not recur or other like solution.A customer service bulletin will be formalized to detail this situation and communicate it to all application and service staff.
|
| |
|
Event or Problem Description
|
|
The x-ray technologist, while transporting portable machine with the retrofit equipment in a backwards motion caught her left foot from the underside of the detector bin of the retrofit unit where it sits close to the floor.This caused an injury to her foot (i.E.Bruising and induced pain).
|
| |
|