The event unit was returned to applied medical.Visual inspection confirmed the complainant's experience of leakage due to fragmentation of an internal rubber component of the seal.Based on the condition of the returned unit and the description of the event, it is likely that the reported event was caused by an instrument.Applied medical's instructions for use states that, "extra care should be used when inserting angular and asymmetrical instruments, such as 'j' hooks and clip appliers.All instruments should be centered axially when inserted through the seal to prevent tearing." applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.This report represents the combined initial and final.Based on the description of the event, the event was deemed not reportable as it was unlikely to cause or contribute to death or serious injury.After evaluation by engineering, the event is deemed reportable due to fragmentation of an internal component of the seal.
|