Medical devices: (b)(4) - nim mainframe response 3.0, serial # (b)(4), lot # 66456200, manufactured date ¿ feb/26/2010, 510(k) # k083124, udi # (b)(4).The nim patient interface (product # 8253200rf) was returned for analysis.Evaluation found that the device had a defective pcba and an incrementing probe failure.Evaluation also found that the unit came in with two incorrect fuses and had a missing o-ring.The device was repaired, missing items were installed, incorrected items were replaced, tested to manufacturer product specifications and returned to the customer.The nim mainframe response (product # 8253001) was returned for analysis.Evaluation could not duplicate reported event of intermittent failure.Evaluation found no fault with the device.The device was tested to manufacturer product specifications and returned to the customer.
|