| Catalog Number |
03P80-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
11/18/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Apoc labeling was evaluated during the investigation as pertaining to the event.The investigation was completed on 04/13/2016.Retain and product was tested and functioning according to specification.Return product was not available for investigation.Investigation: the customer reported obtaining an unexpected k+ result while using 6+ cartridge lot k15261a to test a patient sample.The device history record (dhr) for this lot was reviewed.The lot passed finished goods (fg) release criteria.Forty retained cartridges from the lot were tested using whole blood (wb) and i-stat level 1 control (l1).Testing met the acceptance criteria found in q04.01.003 rev.X, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been identified.Assessment: the complaint investigation concluded there was no product deficiency and that the i-stat cartridge lot is meeting specification.No repeats on the i-stat or laboratory.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2016, abbott point of care was contacted by a customer regarding i-stat cg8+ cartridges that yielded suspected discrepant potassium (k) result on a patient.There was no patient information available at the time of this report.Results from (b)(6) 2015: i-stat k+ - 2.1, antech k+ - 4.3.There was no additional patient information available at the time of this report.Return product is not available for investigation.At the time of the event there was no indication of a product malfunction based on the information provided and assessed by apoc, however this mdr is a retrospective filing in response to an observation from an fda inspection conducted may 8th to 12th, 2017 at abbott point of care ((b)(4)).
|
| |
|
Search Alerts/Recalls
|