MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT B-HCG CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT B-HCG CARTRIDGE Back to Search Results
Catalog Number 05P58-25
Device Problem False Positive Result (1227)
Patient Problem Bronchitis (1752)
Event Date 04/08/2016
Event Type  malfunction  
Manufacturer Narrative
Apoc incident (b)(4).Apoc labeling was evaluated during the investigation as pertaining to the event.The investigation was completed on 05/10/2016.Retain and return product was tested and found to be functioning according to specification.Investigation: the customer reported observing an unexpected result while using b-hcg cartridge lot a15323 to test a patient sample.The device history record for this lot was reviewed.The lot passed finished goods release criteria.Forty retained cartridges from the lot were tested using whole blood.Testing met the acceptance criteria found in q04.01.003 rev.X, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been identified.Assessment: the complaint investigation concluded there was no product deficiency and that the i-stat cartridge lot is meeting specification.
 
Event Description
On (b)(6) 2016, abbott point of care was contacted by a customer regarding i-stat b-hcg cartridges that yielded a suspected false positive result on a female patient who was diagnosed with acute bronchitis.There was no additional patient information available at the time of this report.Time, date of i-stat test, i-stat result: 04/08/16, 12:54, 41.5 iu/l.Time, date of beckman dxi test, lab result: drawn: 12:54,, results: 13:59 <5.0 iu/l.At the time of the event there was no indication of a product malfunction based on the information provided and assessed by apoc; however, this mdr is a retrospective filing in response to an observation from an fda inspection conducted may 8th to 12th, 2017 at abbott point of care ((b)(4)).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT B-HCG CARTRIDGE
Type of Device
B-HCG CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
ottawa, ontario K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key6932300
MDR Text Key90298798
Report Number2245578-2017-00360
Device Sequence Number1
Product Code CHL
UDI-Device Identifier10054749001009
UDI-Public(01)10054749001009
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/09/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2016
Device Catalogue Number05P58-25
Device Lot NumberA15323
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/13/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/19/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-