Catalog Number 06F16-10
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Information (3190)
Date of Event 06/03/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Apoc labeling was evaluated during the investigation as pertaining to the event.The investigation was completed on 10/05/2016.The analyzer was tested and found to be functioning according to specification.Investigation: the device history record for this analyzer was reviewed.The analyzer passed finished goods release criteria.Fifty pt/inr cartridges were tested using level 1 and level 2 controls, and all results were within the acceptance range.The complaint was not reproduced.Assessment: the complaint investigation concluded there was no product deficiency and that the i-stat analyzer is meeting specification.
 
Event or Problem Description
On (b)(6) 2016, abbott point of care was contacted by a customer regarding i-stat pt/inr cartridges that yielded unexpected results on a patient.There was no additional patient information available at the time of this report.(b)(6).At the time of the event there was no indication of a product malfunction based on the information provided and assessed by apoc; however, this mdr is a retrospective filing in response to an observation from an fda inspection conducted may 8th to 12th, 2017 at abbott point of care (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT1 ANALYZER, IMMUNO READY
Common Device Name
I-STAT1 ANALYZER
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key6932451
Report Number2245578-2017-00399
Device Sequence Number13720917
Product Code CGA
UDI-Device Identifier00054749000340
UDI-Public(01)00054749000340
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K001387
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number06F16-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2016
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/13/2017
Initial Report FDA Received Date10/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-