| Catalog Number |
06F16-10 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
06/03/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Apoc labeling was evaluated during the investigation as pertaining to the event.The investigation was completed on 10/05/2016.The analyzer was tested and found to be functioning according to specification.Investigation: the device history record for this analyzer was reviewed.The analyzer passed finished goods release criteria.Fifty pt/inr cartridges were tested using level 1 and level 2 controls, and all results were within the acceptance range.The complaint was not reproduced.Assessment: the complaint investigation concluded there was no product deficiency and that the i-stat analyzer is meeting specification.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2016, abbott point of care was contacted by a customer regarding i-stat pt/inr cartridges that yielded unexpected results on a patient.There was no additional patient information available at the time of this report.(b)(6).At the time of the event there was no indication of a product malfunction based on the information provided and assessed by apoc; however, this mdr is a retrospective filing in response to an observation from an fda inspection conducted may 8th to 12th, 2017 at abbott point of care (b)(4).
|
| |
|