• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT150
Medical Device Problem Codes Fracture (1260); Scratched Material (3020)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Code Available (3191)
Date of Event 09/07/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zxt150 18.0 diopter intraocular lens (iol) was scratched during implantation and an unspecified surgical intervention was required.No additional information was provided.
 
Additional Manufacturer Narrative
Additional information was received and it was learnt that there were no indications of resistance during the intraocular lens (iol) implantation.The lens was fully inserted into the patient's eye and it was being manipulated when the scratch/crack was noted.There was an actual fracture in the center of the iol.Reportedly, the lens was removed and replaced during the same surgical procedure with no further issues.No incision enlargement, no vitrectomy was performed, no sutures were used and no injury was reported.The patient did not have any post-op complications.Date of birth: (b)(6).The surgeon's name was provided: dr (b)(6).(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Additional Manufacturer Narrative
Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection showed that the lens was cut in pieces, most probably to make the removal possible.Considering the condition of the lens additional analysis was not possible.The customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.The dfu states to examine the lens thoroughly to ensure particles have not become attached to it, and examine the lens optical surfaces for other defects.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key6933652
Report Number9614546-2017-00967
Device Sequence Number16429635
Product Code MFK
UDI-Device Identifier05050474605169
UDI-Public(01)05050474605169(17)211213
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date12/13/2021
Device Model NumberZXT150
Device Catalogue NumberZXT150U180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2017
Initial Date Received by Manufacturer 09/13/2017
Supplement Date Received by Manufacturer10/10/2017
01/02/2018
Initial Report FDA Received Date10/09/2017
Supplement Report FDA Received Date11/03/2017
01/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-