• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS MICRO-INTRODUCER KITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS MICRO-INTRODUCER KITS Back to Search Results
Catalog Number H787065072125
Medical Device Problem Code Knotted (1340)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 09/18/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was reported that the disposable device is not available to be returned to the manufacturer for evaluation.An investigation into the root cause of this incident is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.(b)(4).
 
Event or Problem Description
As reported to angiodynamics on september 18, 2017: rn inserted wire into patient and met resistance.When the dr.Tried to withdraw the guidewire and needle, the needle came out, but the wire was stuck.The wire then began to uncoil.The patient was taken to the or to have wire removed.It was reported the wire was successfully removed via cutdown.Upon removal of the guidewire, it was noted the guidewire had knotted.It was reported there was no damage to the vessel due to the event.It was indicated the defective disposable device is not available for return to the manufacturer.
 
Additional Manufacturer Narrative
As the reported defective device was not returned, angiodynamics is unable to perform a device evaluation.However, based on photographs provided by the user, the customer's reported complaint description of guidewire unraveling is confirmed.Although the complaint description is confirmed, without a device evaluation, a definitive root cause for the event cannot be determined.Angiodynamics made the supplier of this device aware of the event via (b)(4).The supplier performed a review of the lot history records, and supplied the statement: "the history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable." a packaging lot history record search was performed and revealed no quality related issues or manufacturing deficiencies at the time of packaging.The instructions for use, which is supplied to the end user with this catalog number contains the following statement "gain percutaneous access with the 21 ga.Entry needle.Advance the 0.018" guidewire through the 21 ga.Needle.Withdraw the entry needle while leaving the 0.018" guidewire in place." a review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO-INTRODUCER KITS
Common Device Name
MICRO-INTRODUCER KITS
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
MDR Report Key6935055
Report Number1319211-2017-00103
Device Sequence Number16495335
Product Code DYB
UDI-Device IdentifierH787065072125
UDI-PublicH787065072125
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K051655
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Nurse
Device Expiration Date05/31/2020
Device Catalogue NumberH787065072125
Device Lot Number5200028
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2017
Supplement Date Received by Manufacturer09/18/2017
Initial Report FDA Received Date10/10/2017
Supplement Report FDA Received Date11/07/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-