| Model Number |
179712740 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
09/07/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is available for evaluation.Investigation will be conducted.Follow up will be filed with findings.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
While performing an extension of an existing spondylodesis, the screw head of an already implanted screw in s1 (left) broke after loosening the innie and rod and retracting the soft tissue at the implant location.This screw was implanted app.2 years ago.It is not clear or unknown whether the screw head was already partially broken before extension surgery.It is therefore believed that the event occurred perioperative.Since the particular vertebral bodies l5-s1 seemed to be fused, no further actions were taken, the screw shaft was left in s1 and the spondy extension was performed as planned.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual examination of the returned device found the screw fractured at the transition zone between shank sphere and shank body.Optical imaging suggests a fatigue fracture.No material defects or other abnormalities have been identified in the fracture analysis report.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.A trend analysis was conducted.No emerging trends were found requiring further actions.A definitive root cause for the polyaxial screw fracturing cannot be positively determined.However, the fracture analysis report suggests a fatigue failure.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|