| Model Number |
MT22495-BLU |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/18/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) 2017, that on (b)(6) 2017, the receiver displayed err 224.No additional patient or event information is available.No product or data was returned for evaluation.The complaint could not be confirmed.A root cause could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The complaint receiver device was returned for evaluation.The device was visually inspected and it passed.The receiver failed to charge and reboot, "call tech support" was displayed on screen.The receiver data was downloaded and shows firmware errors.The reported event of error icon displayed was confirmed.The root cause could not be determined.
|
| |
|
Search Alerts/Recalls
|