| Model Number |
MMT-7008B |
| Health Effect - Clinical Code |
Hypoglycemia (1912)
|
| Date of Event |
10/04/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that they experienced low blood glucose of 50mg/dl.Customer treated low blood glucose with glucose tablets so customer¿s current blood glucose as 130mg/dl.Customer also reports sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 210mg/dl and the sensor glucose was 140mg/dl at the time of the incident.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.The sensor will not returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspected one opened and used sensor and performed continuity resistance test.Sensor passed per specification.Also, performed bicarbonate buffer test.Sensor passed per specifications with accurate readings.See attached file for details.
|
| |
|