Catalog Number 1125350-12
Health Effect - Clinical Code Vessel Or Plaque, Device Embedded In (1204)
Date of Event 09/22/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: a visual inspection was performed on the returned device.The reported stent dislodgement was confirmed.The reported failure to advance the device and the reported difficult to remove were unable to be replicated in a testing environment as they were based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined the reported failure to advance, difficulty removing the device, and stent dislodgement to be related to interactions with the patient heavy tortuosity and heavy calcified anatomy.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported the procedure was to treat a lesion with heavy tortuosity and heavy calcification in the circumflex (cx) coronary artery.Pre-dilatation was performed and the 3.5 x 12 mm xience alpine was advanced, but could not cross the lesion due to the anatomy.During withdrawal, the xience alpine stent dislodged in the left main artery due to the resistance with the anatomy.A balloon catheter was used to compress the dislodged stent against the vessel (healthy tissue) in the left main artery.The patient was in stable condition, but referred for surgery due to the heavily calcified vessel and nothing was able to cross the lesion.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6939143
Report Number2024168-2017-08131
Device Sequence Number872877
Product Code NIQ
UDI-Device Identifier08717648199516
UDI-Public08717648199516
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 10/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/06/2018
Device Catalogue Number1125350-12
Device Lot Number5092841
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2017
Initial Report FDA Received Date10/10/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient Weight81
-
-