• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY RAYNHAM ¿ 1219655 LCS COMP RP AOX INS STD 10; KNEE TIBIAL BEARING/INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY RAYNHAM ¿ 1219655 LCS COMP RP AOX INS STD 10; KNEE TIBIAL BEARING/INSERT Back to Search Results
Catalog Number 129418410
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Pain (1994)
Date of Event 09/12/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Der states that the patient complains of pain on lateral aspect of tibia.Xrays showed tibia tray overhanging.Tray and insert remove and replace with a thicker insert and smaller tray.
 
Additional Manufacturer Narrative
No device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.A worldwide lot specific complaint database search, or device history record (dhr) review was not possible for the unknown lot code(s).Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LCS COMP RP AOX INS STD 10
Common Device Name
KNEE TIBIAL BEARING/INSERT
Manufacturer (Section D)
DEPUY RAYNHAM ¿ 1219655
325 paramount drive
raynham MA 02767
MDR Report Key6939729
Report Number1818910-2017-26322
Device Sequence Number862068
Product Code NJL
UDI-Device Identifier10603295024156
UDI-Public10603295024156
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 09/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number129418410
Device Lot NumberC22161
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/12/2017
Supplement Date Received by Manufacturer10/27/2017
Initial Report FDA Received Date10/11/2017
Supplement Report FDA Received Date11/06/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
-
-