| Catalog Number |
03P78-50 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
02/21/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling was evaluated during the investigation as pertaining to the event.The investigation was completed on 05/24/2017.Retain product was tested and performing to specification.Return product was not available.Investigation: a review of the device history record confirmed the lot passed finished goods release criteria.Retain cartridge testing met the acceptance criteria found in q04.01.003 rev.Z, appendix 1 - product complaint level 2 and level 3 investigation procedure.Assessment: the complaint investigation concluded there was no product deficiency and that the i-stat cartridge lot is meeting specification for ph and pco2.There is not enough information to determine whether an i-stat product malfunction occurred.There are 3 different samples taken by capillary over 14 hours.There were no repeats on any of the 3 samples.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, abbott point of care was contacted by a customer regarding i-stat g3+ cartridges that yielded suspected discrepant ph and pco2 results on a one-month-old male patient in the nicu.No additional patient information was available.Testing performed on 02/21/2017.(b)(6).At the time of the event there was no indication of a product malfunction based on the information provided and assessed by apoc; however, this mdr is a retrospective filing in response to an observation from an fda inspection conducted may 8th to 12th, 2017 at abbott point of care ((b)(4)).
|
| |
|
Search Alerts/Recalls
|