| Model Number |
THD155228E. |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
Death (1802)
|
| Date of Event |
06/17/2017
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated.We are currently waiting for the return of the device sample for evaluation.When the investigation is complete a final report will be filed.
|
| |
|
Event or Problem Description
|
|
Disconnection between the tubing and the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Received a 15.5 titan catheter for evaluation.A visual inspection reveals no damage for defect to the catheter or the luers.The device was forwarded to the device contract manufacturer for evaluation.The investigation of the complaint included a review of the device history record, a review of the returned sample, and the manufacturing process.No rejections or deviations related to the reported failure mode were found during the device history review.There was no visible damaged to the luer.Luer taper was inspected and leak test was performed.Both lumen / luers passed.Root cause could not be determined.No indication the failure was related to the manufacturing process.
|
| |
|