• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET XXL 360 22MM X 50CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET XXL 360 22MM X 50CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0036182250
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/02/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.During the analysis of the returned device, it was revealed that main coil proximal end was within the catheter hub and the main coil was prematurely detached/ separated due to a physical break at the main coil junction.In addition, blood was also noted in the hub.Functional testing could not be performed due to the condition of the returned device.Additional information reported that the device was confirmed to be in good condition prior to use.The presence of blood on the device indicates insufficient flush though this cannot be conclusively determined.Based on the information currently available and the device inspection, it is probable that the coil was damaged during advancement causing the as reported event.An assignable cause of procedural factors will be assigned to the analysed ¿main coil prematurely detached/separated inside patient' as the issue is associated with a product that meets stryker design and manufacture specifications and was used in accordance with the dfu but due to procedural and/or anatomical factors during use, the device performance was limited.
 
Event or Problem Description
During the analysis of the returned device, it was revealed that the main coil was prematurely detached inside the patient.No clinical consequences reported to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XXL 360 22MM X 50CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key6942493
Report Number3008881809-2017-00438
Device Sequence Number1223744
Product Code HCG
UDI-Device Identifier07613327128369
UDI-Public(01)07613327128369(17)191031(10)19380822
Combination Product (Y/N)N
PMA/510(K) Number
K161429
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Model NumberM0036182250
Device Catalogue NumberM0036182250
Device Lot Number19380822
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2017
Initial Report FDA Received Date10/11/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-