LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-80 |
Device Problems
Fluid/Blood Leak (1250); Device Operational Issue (2914)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/19/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The heater-cooler device 16-02-80 is not distributed in the usa.However, it is similar to the heater-cooler system 3t device 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was informed by the customer that a third party service provider identified a faulty compressor to be the root cause of the reported event.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
|
Event Description
|
Livanova (b)(4) received a report that a heater-cooler system 3t failed during priming due to earth leakage of the compressor caused as a result of a leak in the device water tanks.There was no patient involvement.
|
|
Manufacturer Narrative
|
Through further follow-up communication with the customer, livanova learned that the hospital performed the investigation and determined that the fault was caused by a leakage in one of the watertanks.Based on internal livanova investigation, it was determined that in specific conditions a hole can be generated in a defined area of the evaporator located in the patient tank.Such hole might create a leak of refrigerant in the tank and consequent water ingress in the cooling circuit.Such situation would then damage the compressor.The hospital declined to send the device back for further investigation.Therefore, no further investigation is possible.Corrective actions are in progress for this issue.
|
|
Search Alerts/Recalls
|
|
|