LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problems Fluid/Blood Leak (1250); Device Operational Issue (2914)
Patient Problem No Patient Involvement (2645)
Event Date 09/19/2017
Event Type  malfunction  
Manufacturer Narrative
The heater-cooler device 16-02-80 is not distributed in the usa.However, it is similar to the heater-cooler system 3t device 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was informed by the customer that a third party service provider identified a faulty compressor to be the root cause of the reported event.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t failed during priming due to earth leakage of the compressor caused as a result of a leak in the device water tanks.There was no patient involvement.
 
Manufacturer Narrative
Through further follow-up communication with the customer, livanova learned that the hospital performed the investigation and determined that the fault was caused by a leakage in one of the watertanks.Based on internal livanova investigation, it was determined that in specific conditions a hole can be generated in a defined area of the evaporator located in the patient tank.Such hole might create a leak of refrigerant in the tank and consequent water ingress in the cooling circuit.Such situation would then damage the compressor.The hospital declined to send the device back for further investigation.Therefore, no further investigation is possible.Corrective actions are in progress for this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6942875
MDR Text Key90312316
Report Number9611109-2017-00805
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/26/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/02/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-