MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL CASHMERE 14 - CERECYTE MICROCOIL; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL CASHMERE 14 - CERECYTE MICROCOIL; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Catalog Number CRC14082030
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/22/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Mfg name codman & shurtleff, inc.Dba depuy synthes products, inc.Three attempts to obtain additional information were unsuccessful.It was reported that the device would be returned for analysis; however, the device has not yet been returned.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported by a healthcare professional, during the procedure, the cashmere coil (crc14082030/c27073) could not be detached.They used another one to complete the procedure.There was no report of patient injury.It was reported that the device would be returned for analysis.
 
Additional Manufacturer Narrative
Multiple attempts to obtain the device for analysis and obtain additional information were unsuccessful.If additional information or product returned at a later date, a supplemental mdr will be submitted.Conclusion: as reported by a healthcare professional, during the procedure, the cashmere coil (crc14082030/c27073) could not be detached.They used another one to complete the procedure.There was no report of patient injury.It was reported that the device would be returned for analysis; however, after multiple requests, the device was not returned and no additional information provided.The device was not available for investigation.A review of manufacturing documentation associated with this lot presented no issues during the manufacturing or inspection processes related to the reported complaint.The failure to detach could not be confirmed without product return for analysis.The root cause of the event could not be determined.There is no current safety signal identified related to the reported event based on review of complaint history for the device.Since there was no evidence to suggest the event was related to a manufacturing issue, no corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CASHMERE 14 - CERECYTE MICROCOIL
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
rue girardet 29
le locle jura
SZ 
MDR Report Key6947128
Report Number1226348-2017-00169
Device Sequence Number16330005
Product Code HCG
UDI-Device Identifier00878528004646
UDI-Public(01)00878528004646(17)190405(10)C27073
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K072173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 09/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/05/2019
Device Catalogue NumberCRC14082030
Device Lot NumberC27073
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/20/2017
Supplement Date Received by Manufacturer11/29/2017
Initial Report FDA Received Date10/13/2017
Supplement Report FDA Received Date11/30/2017
Patient Sequence Number1
-
-