MAUDE Adverse Event Report: UNITED STATES ENDOSCOPY GROUP, INC. BIOGUARD AIR/WATER & SUCTION VALVE KIT; BIOPSY VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNITED STATES ENDOSCOPY GROUP, INC. BIOGUARD AIR/WATER & SUCTION VALVE KIT; BIOPSY VALVE Back to Search Results
Model Number 00711784
Health Effect - Clinical Code Exposure to Body Fluids (1745)
Date of Event 09/27/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was not returned to us endoscopy.The lot history record was reviewed and found no issues during manufacture; the device was built to specification.The instructions for use contain the following warnings and precautions: "exposure to bodily fluids may occur during connection or disconnection of these devices; adherence to body substance isolation protocols is the responsibility of the user.Do not leave a device hanging from the valve.Doing so can cause the creation of a larger valve slit/hole that may cause leakage.If the lid of the valve is opened while attached to the endoscope during a procedure, scope suction will be compromised and leakage may occur.If leakage occurs, a sterile gauze should be used to cover the valve." in-service training has been provided to the user.Not returned to manufacturer.
 
Event or Problem Description
The bioshield® biopsy valve is used to cover the opening to the biopsy/suction channel inlet of a gastrointestinal endoscope.The bioshield® biopsy valve provides access for endoscopic device passage and exchange, helps maintain insufflation and minimizes leakage of biomaterial from the biopsy port throughout the gastrointestinal endoscopic procedure.The user facility reported the biopsy valve in the bioguard air/water and suction valve kit sprayed fluid that contacted a physician's and a nurse's personal protective equipment.There is no report of patient or user harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOGUARD AIR/WATER & SUCTION VALVE KIT
Common Device Name
BIOPSY VALVE
Manufacturer (Section D)
UNITED STATES ENDOSCOPY GROUP, INC.
5976 heisley rd
mentor OH 44060
MDR Report Key6948829
Report Number1528319-2017-00029
Device Sequence Number1142082
Product Code ODC
UDI-Device Identifier10816765013004
UDI-Public(01)10816765013004(17)190802(11)170803(10)1708046
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K151345
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 10/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date08/02/2019
Device Model Number00711784
Device Catalogue Number00711784
Device Lot Number1708046
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2017
Initial Report FDA Received Date10/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-