• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN SILVERHAWK BTK; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN SILVERHAWK BTK; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Catalog Number P4044
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/14/2017
Type of Reportable Event Malfunction
Event or Problem Description
The physician was attempting to use one silverhawk directional atheterctomy device to treat a slightly tortuous and moderately calci fied lesion located in the tibial/popliteal trunk.The device was removed from its packaging and inspected with no issues noted.The device was prepped per the ifu.It was reported that the cutter window jammed during the procedure and that the physician was unable to close the window during cutting.The physician was unable to return the cutter to the housing before the device was removed from the patient.The physician removed the device and used a new device to complete the procedure.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILVERHAWK BTK
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
COVIDIEN
9775 toledo way
irvice CA 92618
MDR Report Key6950203
Report Number2183870-2017-00448
Device Sequence Number10987819
Product Code MCW
UDI-Device Identifier00821684074107
UDI-Public00821684074107
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K061188
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/20/2019
Device Catalogue NumberP4044
Device Lot NumberA448024
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2017
Initial Report FDA Received Date10/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-