| Model Number |
ZMA00 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017.(b)(4).All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zma00 13.5 diopter intraocular lens (iol) was implanted (b)(6) 2017.It was later explanted on (b)(6) 2017 because the lens would not stay centered.The lens was replaced with the same model, but different diopter of 14.0.Reportedly, the patient is doing well with no post-operative patient injury.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation?: yes, returned to manufacturer on 10/23/2017.Device returned to manufacturer?: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Search Alerts/Recalls
|