• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PUMP MMT-1714K 630G BLACK MMOL CANADA; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PUMP MMT-1714K 630G BLACK MMOL CANADA; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1714K
Device Problems No Display/Image (1183); Moisture Damage (1405); Device Inoperable (1663); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/10/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump was exposed to moisture and had blank display.The customer¿s blood glucose was 93.6 mg/dl at the time of the incident.The customer reported that he was swimming with the insulin pump.The customer reported that he noticed that the display was blank.Troubleshooting was not performed at the time of call because the customer reported that the insulin pump was not working.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Unit received with blank display due to corroded electronic assemblies.Unit received with corroded battery tube and corroded motor home switch.Unit received with minor scratches on case, cracked case corner of the belt clip rails near the battery compartment, cracked retainer, stained serial number label and minor scratches on lcd window.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP MMT-1714K 630G BLACK MMOL CANADA
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key6953614
MDR Text Key89838700
Report Number2032227-2017-59058
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169782396
UDI-Public(01)00643169782396
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1714K
Device Catalogue NumberMMT-1714K
Device Lot NumberHG1T8Q2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/10/2017
Initial Date FDA Received10/16/2017
Supplement Dates Manufacturer Received10/10/2017
Supplement Dates FDA Received03/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-