| Catalog Number |
826-213 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/22/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, the cable broke.The product came in contact with the patient.No fragments remained in the patient.No patient complications were reported as a result of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The patient underwent hip replacement surgery due to congenital hip dysplasia.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: the product was returned with four distinctive failures in the cable.Under microscopic review three of the failures show necking at the ends of the individual strands and a tightening of the original cable twist.These observations are consistent with an overload from tension.The other failure appears to be the result of a cut.There is a uniformity to the angle of the single strands and witness marks can be seen on the face of the strands.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|