MEDTECH SA ROSA SURGICAL DEVICE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA 2.5.8
Device Problem Misconnection (1399)
Patient Problem No Patient Involvement (2645)
Event Date 09/22/2017
Event Type  malfunction  
Manufacturer Narrative
The device has not been evaluated yet for investigation purpose.Once the evaluation will be performed, a follow-up medwatch report will be submitted.
 
Event Description
It was reported that the robot seems turned on because the controller made some noise but nothing was displayed on the screen.After a call to the field service engineer the surgeon checked the electrical contact of the screen cables and the screen then functioned normally.
 
Manufacturer Narrative
According to results of the investigation, the touch screen support arm was modified to be in three sections instead of two sections.Therefore the screen cable was marginally too short and that when the touch screen was being adjusted or fully extended the cable was losing contact.This is what caused the reported screen failures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA SURGICAL DEVICE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key6954007
MDR Text Key90468331
Report Number3009185973-2017-00736
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
PK101791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Repair
Type of Report Initial,Followup
Report Date 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROSA 2.5.8
Device Catalogue NumberN/A
Device Lot NumberN/A
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/18/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/11/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-