Model Number 53316
Health Effect - Clinical Code Abrasion (1689)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the drainage eyes on the catheter had sharp edges.Allegedly, the patient experienced scraping on the urethra; however, no medical intervention was required.
 
Additional Manufacturer Narrative
Received 1 used intermittent catheter with package label.The reported event was unconfirmed.The sample was visually inspected for holes or damage on the package, legibility of printed numbers and letters, expiration date, correct impression, print color, missing parts of printing, position of printing, damaged component, and foreign matter.The drainage eyes did not present a sharp edge on its tip.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "for urological use only.Urinary catheters are intended for use for bladder management including urine drainage, collection and measurement.The devices are passed to the urinary bladder via the urethra.Warnings: this is a single use device.Do not reuse.Reuse of a single use device increases the risk of catheter acquired urinary tract infections.How to prepare and use your urinary catheter: wash your hands thoroughly with soap and water.Peel open the pack at the funnel end to expose approximately 2¿- 4¿ of the catheter.Don¿t remove the catheter yet.Wash the area around the meatus before catheterizing.Wash your hands again.Using the catheter: with sure-grip¿ sleeve: hold the sure-grip¿ sleeve with your dominant hand and squeeze it to hold the catheter shaft as you remove the catheter from the pack.Next, hold the catheter funnel above the sure-grip¿ sleeve with your other hand and slide the sure-grip¿ sleeve down the shaft, stopping at about 6¿ from the tip.Release the funnel.Using the sure-grip¿ sleeve to hold the catheter firmly, lubricate catheter tip with lubricating gel.While gripping the catheter with the sure-grip¿ sleeve, gently pass the tip of the catheter into your urethra until the sure-grip¿ sleeve nears the meatus.If there is no urine flow, loosen the tension on the sure-grip¿ sleeve to allow it to slide back towards the funnel end.Re-grip sure-grip¿ sleeve and continue to insert the catheter into the urethra.Repeat until urine starts to flow.Without sure-grip¿ sleeve: remove the catheter from the pack.Lubricate the catheter tip with lubricating gel.Gently pass the tip of the catheter into your urethra and advance the catheter until urine starts to flow.Try to keep the catheter steady until urine stops flowing.When urine stops flowing, slowly withdraw the catheter, stopping if flow starts again, until the last few drops have drained.Finish by disposing of the catheter and its packaging.Wash your hands with soap and water." (b)(4).
 
Event or Problem Description
It was reported that the drainage eyes on the catheter had sharp edges.Allegedly, the patient experienced scraping on the urethra; however, no medical intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGIC3¿ INTERMITTENT CATHETER
Common Device Name
MAGIC 3 CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key6954220
Report Number1018233-2017-05298
Device Sequence Number795202
Product Code KOD
UDI-Device Identifier00801741068430
UDI-Public(01)00801741068430
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K000723
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 01/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/28/2022
Device Model Number53316
Device Catalogue Number53316
Device Lot NumberNGBQ1037
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2017
Initial Date Received by Manufacturer 09/28/2017
Supplement Date Received by Manufacturer01/03/2018
Initial Report FDA Received Date10/17/2017
Supplement Report FDA Received Date01/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-