| Model Number |
482317535 |
| Medical Device Problem Codes |
Material Deformation (2976); Material Integrity Problem (2978)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
| Date of Event |
09/19/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; when the screw was removed during surgery, the tulip head was stuck.Therefore, the screw was removed with kochel.Update (b)(6): when the surgeon was performed final tightening of screw, the anti torque key came off from the screw and the screw hole thread was broken.Therefore, the screw was removed.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.Visual inspection showed the threads were found to be deformed which can be indicative of blocker misalignment or cross threading.Conclusion: the plausible root cause of the event is determined to be use error- misalignment of the blocker and the tulip.
|
| |
|
Event or Problem Description
|
|
It was reported that; when the screw was removed during surgery, the tulip head was stucked.Therefore, the screw was removed with kochel.Update 10/16/2017: when the surgeon was performed final tightening of screw, the anti torque key came off from the screw and the screw hole thread was broken.Therefore, the screw was removed.
|
| |
|