MAUDE Adverse Event Report: MDT PUERTO RICO OPERATIONS CO SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT PUERTO RICO OPERATIONS CO SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problems Electromagnetic Interference (1194); Mechanical Jam (2983)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/16/2017
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider via a company representative regarding a patient receiving morphine 10mg/ml at 2.7mg/day via an implantable pump.The indication for use was non-malignant pain.It was reported a motor stall was seen at initial interrogation.The patient recently had an mri.The logs were read and a motor stall recovery had not yet occurred.It had not been less than 2 hours since the patient had exited the mri field.Telemetry state was reviewed and re-interrogating the pump with logs.There were no patient symptoms and no further patient complications reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received reported that they re-read the pump 30 minutes and the pump had recovered from the motor stall.All was good.No further patient complications were reported or anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer (Section G)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6954880
MDR Text Key90550134
Report Number3004209178-2017-21936
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00613994779229
UDI-Public00613994779229
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 10/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2014
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/16/2017
Initial Date FDA Received10/17/2017
Supplement Dates Manufacturer Received10/17/2017
Supplement Dates FDA Received10/18/2017
Date Device Manufactured08/04/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0591-2009
Patient Sequence Number1
Patient Age68 YR
Patient Weight68
-
-