MAUDE Adverse Event Report: AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012445-15
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/04/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: the device was returned.A visual and dimensional inspection was performed on the returned device.The reported kinked shaft and shaft detachment were confirmed.The reported failure to advance could not be tested as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined that the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported the procedure was performed to treat a lesion in the obtuse marginal vessel with moderate calcification.The 2.0 x 15 mm nc trek balloon dilatation catheter (bdc) was advanced for dilatation.During advancement of the nc trek, resistance was met with the calcification and the proximal shaft kinked, then separated outside the anatomy.This was prior to reaching the lesion.The nc trek was removed without issue and a non-abbott bdc was used for successful dilatation.A 3.0 x 8 mm xience alpine stent was successfully deployed to treat the lesion.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6955472
Report Number2024168-2017-08262
Device Sequence Number1142125
Product Code LOX
UDI-Device Identifier08717648151651
UDI-Public08717648151651
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K110134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 10/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2018
Device Catalogue Number1012445-15
Device Lot Number5072761
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2017
Initial Report FDA Received Date10/17/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age79 YR
Patient Weight82
-
-