Model Number AC4330-01
Health Effect - Clinical Code No Information (3190)
Date of Event 10/13/2017
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the metal hook of the saber 30 wand broke off inside the patient.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.A relationship, if any, between the subject device and the reported event could not be determined since the product was not returned for evaluation.Visual inspection and functional testing could not be performed because the device in question was not returned for evaluation.Thus, the complaint could not be verified, nor could a root cause be determined with confidence.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.Factors not associated with the manufacture or design of the device which could result in damage to the tip could be the device coming into abrupt contact with a hard (metallic) object, improper insertion or removal from an apparatus or excessive force applied to the tip can also lead to unintended detachment.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABER 30 DEGREE ICW
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key6955821
Report Number3006524618-2017-00318
Device Sequence Number16466478
Product Code GEI
UDI-Device Identifier00817470004854
UDI-Public(01)00817470004854(17)200430(10)1192419
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K033584
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Model NumberAC4330-01
Device Catalogue NumberAC4330-01
Device Lot Number1192419
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/13/2017
Supplement Date Received by Manufacturer10/13/2017
Initial Report FDA Received Date10/17/2017
Supplement Report FDA Received Date12/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-