| Model Number |
E SERIES |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The malfunction was observed during incoming testing; however after disassembling the device to determine root cause, the malfunction was no longer seen.The device was put through extensive testing without duplicating the malfunction.The control to display power cable was replaced as a precaution.The device passed the final test procedure, was recertified, and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device powered on without display or the display intermittently flickered and no clinical data could be seen.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|