| Model Number |
ZCT525 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but the best estimate date is during (b)(6) 2017.Explant date: if explanted, give date: not applicable, as the lens remains implanted.(b)(4).All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zct525 22.5 diopter intraocular lens (iol) was implanted in the right eye (od) on (b)(6) 2017 and rotated to 135 degrees from the target degree of 100, which was a difference of 35 degrees.In addition, the patient's post-operative visual acuity was 20/70 and she complained of blurry vision.Therefore, a second surgery was performed on (b)(6) 2017 to rotate the lens back to the 100 degree target.Post-operatively, the lens moved to 110 degrees and the patient's visual acuity was 20/30.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product was not returned to the manufacturing site as it remains implanted.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.S has been submitted.
|
| |
|
Search Alerts/Recalls
|