MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER; BARD X-FORSE NEPHROSTOMY BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER; BARD X-FORSE NEPHROSTOMY BALLOON Back to Search Results
Catalog Number 995101
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/08/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that the sheath was deformed upon opening the package.
 
Additional Manufacturer Narrative
The reported event is confirmed.Visual inspection of the sample showed the unit was deformed on one end.The sheath was partially flattened instead of round.No pieces were fractured or missing.The deformation of the sheath appeared to have been ductile rather than brittle, indicating that strain on the unit was applied slowly.This type of slow strain can occur during shipping, handling, or storage.A dhr could not be performed as no lot number was provided from the packaging of the unit.The root cause of the event could not be determined.According to the fmea, a potential root cause of the reported event is the support tube packaging for the sheath did not provide support as intended, either due to inadequate design or errors made in the packaging process.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: "caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Description the x-force® nephrostomy balloon dilation catheter is a dual lumen catheter with a 24 (8mm) or 30fr (10mm) balloon mounted on the distal tip.It has a radiopaque tip and a radiopaque marker beneath the balloon.The lumen labeled with rated burst pressure (xx atm) is for balloon inflation.The other lumen allows the catheter to track over a 0.038" (.97mm) diameter guidewire and can be used for monitoring of pressure or the infusion of medication and/or contrast medium.Each balloon inflates to a stated diameter and length at a specific pressure ¿ typically at 10 atm.The balloon dilation catheter comes packaged with a refolding tool and a working sheath.It is available with or without an inflation device.It comes sterile and is for single patient use only.Indications for use the x-force® nephrostomy balloon dilation catheter is recommended for use in the dilation of the nephrostomy tract and for placement of the working sheath.Contraindications do not use the x-force® nephrostomy balloon dilation catheter in the presence of conditions which create unacceptable risk during the dilation of the nephrostomy tract.Warnings: ¿ if resistance is felt when removing either the catheter or the guidewire from the working sheath, stop and consider removing them as a single unit to prevent damage to the product.Applying excessive force to the catheter can result in tip breakage or balloon separation.¿ do not use air or any gaseous substances as a balloon inflation media, always use sterile liquid media.¿ this is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Precautions: ¿ only a physician who has an understanding of the clinical applications, technical principles and associated risks associated with balloon dilation of the nephrostomy tract should use this device.¿ after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and with applicable laws and regulations.Potential complications the complications that may arise from a balloon dilation procedure include tissue trauma and perforation.Inspection prior to use the x-force® nephrostomy balloon dilation catheter is a sterile, single patient use device.Carefully inspect the catheter and the sterile packaging for signs of damage that may have occurred during shipment.Do not use the product if damage is evident.Preparation of the catheter all x-force® nephrostomy balloon dilation catheters contain air in the balloon lumen.The air must be removed to allow liquid to fill the balloon when it is inflated.Remove the protective sheath from the balloon.Attach the inflation device to the connector on the balloon lumen.Open the stopcock, and draw back on the inflation device to remove the air from the balloon catheter close the stopcock, remove the inflation device, depress the plunger to remove any air and reattach to the balloon catheter.Repeat steps 3-4 until all air is removed from the balloon lumen.Catheter insertion prior to insertion, place the working sheath over the balloon and position it proximal to the balloon.Note: dilation procedures should be conducted under fluoroscopic guidance with appropriate x-ray equipment or direct vision.Introduce the catheter carefully over a 0.038" (.97mm) guidewire and place it in the area that needs to be dilated.Use the radiopaque marker to aid in proper positioning.Caution: do not advance or withdraw the catheter or guidewire against any significant resistance.The cause of the resistance must be determined fluoroscopically and remedial action taken.Inflating the balloon catheter fill the inflation device (eagle¿ inflation device) with sterile, liquid media.Attach the inflation device to the balloon lumen.Open the stopcock and inflate the balloon.Once dilation has been attained, advance the sheath over the balloon.Warning: do not use air or any gaseous substances as a balloon inflation media, always use sterile liquid media.Note: the use of an inflation device with a pressure gauge is highly recommended to make sure adequate pressure is applied and the maximum limits of the balloon are not exceeded.Note: slight reduction in balloon pressure may be required if resistance is encountered when advancing the sheath over the balloon.Caution: do not exceed the recommended rated burst pressure (rbp) for this device.Balloon rupture may occur if the rbp rating is exceeded.Please refer to the device label for the recommended rbp.Deflating the balloon catheter deflate the balloon using an inflation device.Since deflation times vary based on balloon sizes and lengths, check fluoroscopically to confirm deflation before attempting to withdraw.Attach the inflation device (eagle¿ inflation device) and remove the solution from the balloon by pulling back on the inflation device.Remove the inflation device from the balloon catheter.This will allow ambient pressure to enter, relaxing the balloon.Gently withdraw the catheter.Use of a gentle counterclockwise twisting motion is recommended when removing the catheter.Warning: if resistance is felt when removing either the catheter or the guidewire from the working sheath, stop and consider removing them as a single unit to prevent damage to product.Applying excessive force to the catheter can result in tip breakage or balloon separation.Using the refolding tool the balloon catheter is supplied with a refolding tool.This aids in preparing the balloon for another insertion during the procedure.To use: manually compress the balloon.Position the refolding tool at one end.Twist the refolding tool counter clockwise and push down on the balloon until it goes the entire length of the balloon.Once the balloon is folded, remove the refolding tool and proceed with the procedure.Storage store in a cool, dry and dark place.Do not store near radiation or ultraviolet light sources as these may damage product materials." (b)(4).
 
Event or Problem Description
It was reported that the sheath was deformed upon opening the package.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER
Common Device Name
BARD X-FORSE NEPHROSTOMY BALLOON
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key6957879
Report Number1018233-2017-05341
Device Sequence Number719190
Product Code LJE
UDI-Device Identifier00801741080722
UDI-Public(01)00801741080722
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K063632
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number995101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2017
Initial Date Received by Manufacturer 09/29/2017
Supplement Date Received by Manufacturer12/28/2017
Initial Report FDA Received Date10/18/2017
Supplement Report FDA Received Date12/31/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-