| Catalog Number |
AR-1380TB |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
08/31/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.Device history record review revealed nothing relevant to this event.No device malfunction can be identified.At this time, it cannot be determined how the device may have caused or contributed to the incident.An evaluation of the device cannot be performed as the device was discarded by the facility.Additional information has been requested but not made available.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.This is the first complaint of this type for this part/lot combination.The potential causes of this event are being communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was originally reported that during the surgery the screw broke.Investigation update: one half remained inside the patient the other half was retrieved.The screw does not enter easily into the bone, but the surgeon tried anyway (with force) and so the screw broke on insertion.
|
| |
|