Model Number 2002
Medical Device Problem Codes Failure to Charge (1085); Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/29/2017
Type of Reportable Event Malfunction
Event or Problem Description
This patient was implanted with the maestro rechargeable system on (b)(6) 2017.This subject has reported issues charging the implanted rechargeable neuroregulator.Using a charged mobile charger, communication required for charging could not be maintained.Upon device interrogation, no alarms or error codes were noted.An alarm clearing sequence was run during a clinic visit on (b)(6) 2017.Following this, a charge was successfully initiated.Patient left the clinic receiving therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAESTRO RECHARGEABLE SYSTEM
Common Device Name
RECHARGEABLE NEUROREGULATOR
Manufacturer (Section D)
ENTEROMEDICS, INC.
2800 patton road
saint paul MN 55113
MDR Report Key6958720
Report Number3005025697-2017-00023
Device Sequence Number795215
Product Code PIM
UDI-Device Identifier00857334004262
UDI-Public00857334004262
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P130019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date08/10/2017
Device Model Number2002
Device Catalogue Number2002
Device Lot Number093G00716
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2017
Initial Report FDA Received Date10/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age47 YR
Patient Weight113
-
-