MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-522NAL
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 10/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was 92 mg/dl at the time of the incident.The customer also reported that the insulin was squirting during manual prime process.The customer stated that the drive support cap was flush.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump is expected to be returned for analysis.
 
Additional Manufacturer Narrative
Pump passed the displacement test.Unit received with motor error alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform prime test, excessive no delivery test and occlusion test or verify the compromised force sensor system alarm due to motor error alarm.Pump received with minor scratched lcd window, cracked case at the display window corners, cracked lcd window, cracked belt clip slot, cracked battery tube threads, missing end cap sticker, stained address/serial number label and cracked reservoir tube lip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key6958875
Report Number2032227-2017-59539
Device Sequence Number1142432
Product Code OYC
UDI-Device Identifier00613994402110
UDI-Public(01)00613994402110
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 04/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-522NAL
Device Catalogue NumberMMT-522NAL
Device Lot NumberA5522NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2017
Initial Date Received by Manufacturer 10/12/2017
Supplement Date Received by Manufacturer03/26/2018
Initial Report FDA Received Date10/18/2017
Supplement Report FDA Received Date04/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age40 YR
Patient Weight178
-
-