The device was returned to conmed for evaluation.Visual inspection found the lower jaw broken with no other signs of damage.The mechanism of the device functioned as expected and there was no noticeable bend in the device's shaft.This device is a technique dependent device and the most likely cause of this suspected failure is user related or degradation over time.A review of the manufacturing documents was unable to be completed as the manufacturing documents are held by the manufacture classic wire cut co.(b)(4).The instructions for use advise the user of the following.Avoid mechanical shock or over stressing the instrument which may shorten the life of the instrument.Do not use if parts are broken, cracked or worn, or device function may be compromised.Prior to use, inspect instrument to ensure it is in good physical condition and functions properly.There should be no loose, broken or misaligned parts.To review the reported type of failure, an investigation has been initiated.This issue will continue to be monitored through the complaint system to assure patient safety.
|