| Model Number |
VLOCL1413 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/29/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the procedure, the loop broke while suturing.Nothing fell into the patient cavity and the procedure was completed with another device.There was no injury caused to the patient and no medical intervention was required as a result of the device issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary post market vigilance (pmv) led an evaluation of one device.The visual inspection of the sample noted 1 suture and 1 needle.Microscopic inspection of the loop noted material transfer which is an indication that the loop was welded.This serves as evidence that the loop was broken under tension; however, since the sample was received with the loop open, the force required to break the loop cannot be determined.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during an orthopedic procedure, the loop broke while suturing.Nothing fell into the patient cavity and the procedure was completed with another device.There was no injury caused to the patient and no medical intervention was required as a result of the device issue.
|
| |
|
Search Alerts/Recalls
|