BARD ACCESS SYSTEMS DOT PG PRO 20G 8CM FULL; INTRAVASCULAR CATHETER Back to Search Results
Model Number F120081
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of rebs1202 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
Per sales rep, the facility reported a pin hole was found in the power pro midline catheter.The catheter was removed.No reported patient injury.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a hole in the catheter was confirmed and was determined to be related to use of the device.One 20g x 8cm powerglide catheter, which consisted of the purple hub and pink strain relief sleeve, was returned for investigation.The deployment system was not returned for investigation.A functional test revealed a hole in the catheter 6mm from the distal end of the pink sleeve.A microscopic examination revealed that the catheter was scored longitudinally with a circumferential slit at the distal end of the longitudinal score line.Sharp edges were observed along the slit with striations on adjoining surfaces of the breach in the tubing.These characteristics at the leak site are consistent with a needle puncture.The product ifu warns, ¿once the catheter has been advanced, do not re-insert the needle back into the catheter or pull the catheter back onto the needle.This may result in damage to the catheter.If the catheter needs to be repositioned, either do so without the aid of the needle, or remove both the catheter and the needle as a unit to prevent the needle from damaging or shearing the catheter.¿ the ifu also states, ¿do not perforate, tear, or fracture the catheter with the needle or guidewire during the procedure.¿ a lot history review (lhr) of rebs1202 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
Per sales rep, the facility reported a pin hole was found in the power pro midline catheter.The catheter was removed.No reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT PG PRO 20G 8CM FULL
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key6961510
Report Number3006260740-2017-01855
Device Sequence Number13476168
Product Code FOZ
UDI-Device Identifier00801741110573
UDI-Public(01)00801741110573
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K153280
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 01/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberF120081
Device Catalogue NumberF120081
Device Lot NumberREBS1202
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2017
Event Location Hospital
Initial Date Received by Manufacturer 10/17/2017
Supplement Date Received by Manufacturer11/13/2017
Initial Report FDA Received Date10/19/2017
Supplement Report FDA Received Date01/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-