| Medical Device Problem Codes |
Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the reporter contacted animas, alleging a call service alarm (call service alarm issue) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1: date of submission 22-dec-2017 - device evaluation: the device has been returned and evaluated by product analysis on 29-nov-2017 with the following findings: during investigation, a ¿cs069¿ alarm was reproduced on power up.The pump was unable to recover from the ¿cs069¿ after reboot.The ¿cs069¿ failure was defined as a language file corruption while retrieving the language text.The ¿cs069¿ failure was written as a ¿cs087¿ failure in the black box.The error was resident to the flash chip (u39).Unrelated to the original complaint, the battery compartment was found to be cracked.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
| |
|