The returned valve was patent.Therefore the conditions of the complaint could not be duplicated by laboratory personnel.It met the requirements for reflux, and leak testing.However, the valve did not meet requirements for pressure-flow, and pre-implantation testing.There was crystalline debris noted within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open affecting the flow of fluid through the valve and may result in fluid reflux.The instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ all valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|