• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY CSF-FLOW CONTROL VALVE, BURR HOLE, 12 MM, MEDIUM PRESSURE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY CSF-FLOW CONTROL VALVE, BURR HOLE, 12 MM, MEDIUM PRESSURE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 42534
Medical Device Problem Code Inaccurate Flow Rate (1249)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/20/2017
Type of Reportable Event Malfunction
Event or Problem Description
It was reported there was no flow through the valve when the nurse did the testing prior to implant.The valve was replaced and there was no injury to the patient.
 
Additional Manufacturer Narrative
The returned valve was patent.It met the requirements for reflux, pressure-flow, preimplantation, and leak testing.Therefore the conditions of the complaint could not be duplicated by laboratory personnel.The instructions for use that accompany the device instruct that when performing the patency check: a.Place the inlet connector of the valve into sterile isotonic saline.B.Depress the valve dome.C.Place a finger over the opening at the end of the outlet connector.D.Release the depressed dome.If fluid enters the reservoir, the inlet connector and flow control membrane valve are patent.* e.Remove finger from the outlet connector opening.F.Depress dome.If fluid flows out of the outlet connector, the valve is patent.* * note: it may be necessary to depress the valve dome more than once to initiate flow, especially if the ventricular catheter has been attached prior to patency testing.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSF-FLOW CONTROL VALVE, BURR HOLE, 12 MM, MEDIUM PRESSURE
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key6964482
Report Number2021898-2017-00526
Device Sequence Number19213173
Product Code JXG
UDI-Device Identifier00643169466067
UDI-Public00643169466067
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K831678
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 11/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2020
Device Model Number42534
Device Catalogue Number42534
Device Lot NumberE00783
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2017
Supplement Date Received by Manufacturer10/27/2017
Initial Report FDA Received Date10/20/2017
Supplement Report FDA Received Date11/21/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-