| Model Number |
3889-28 |
| Medical Device Problem Codes |
Unsealed Device Packaging (1444); Device Damaged Prior to Receipt by User (1570); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/18/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a manufacturer representative (rep) regarding a patient who was being implanted with a neurostimulator.It was reported that the sterile packaging of the lead looked compromised.Although it was not opened, it appeared there was slight water damage on the tyvek component of the packaging.The healthcare professional (hcp) made the decision to use the lead and, once it was deployed in the patient, they tested the electrodes looking for patient sensory and motor response.The 3 and 2 electrodes seemed to work and produced the appropriate responses, but the 1 and 0 electrodes did not garner a response from the patient.A decision was made to open a new lead and the problem was resolved.There were no symptoms or further complications reported or anticipated.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of product id# 3889-28 found no anomaly as the returned device passed all functional testing in the laboratory.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of medical product: product id: neu_stylet_acc, serial# unknown, product type: accessory.Analysis results were not available at the time of this report.A follow-up report will be sent when analysis is completed.If information is provided in the future, a supplemental report will be issued.
|
| |
|