Model Number S7
Device Problem Imprecision (1307)
Patient Problems Injury (2348); Iatrogenic Source (2498)
Event Date 09/25/2017
Event Type  Injury  
Manufacturer Narrative
No parts have been received by the manufacturer for evaluation.Part not returned for analysis.
 
Event Description
A medtronic representative reported that, while in a lead placement deep brain stimulation (dbs) procedure, an imprecision of 5mm was observed.The representative reported that an entry point was determined with a medtronic imaging system image acquisition.The burr hole and nexframe were then attached and a second imaging system image acquisition found that the plan and actual location varied.The entry point was reset and the navigation system displayed critical structures would be compromised.The surgeon indicated visually that no critical structures were present.The surgeon then placed the lead and a confirmation image acquisition found that the lead was placed 3mm parallel to the plan and recordings were not great.The surgeon then removed the lead, re-aligned to their target and performed a partial placement.Following partial placement, an image acquisition was performed to confirm trajectory.The lead was then placed and a confirmation image acquisition found accurate placement.No additional information was provided.
 
Manufacturer Narrative
Additional information: there was a reported delay to the procedure of thirty minutes due to this issue.
 
Manufacturer Narrative
The software investigation found that a probable cause was unable to be determined without further information since the on-going investigation proved to be inconclusive based on the information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key6966725
MDR Text Key89919237
Report Number1723170-2017-04277
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 12/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? No
Date Manufacturer Received11/17/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/27/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age72 YR
Patient Weight76
-
-